Metrohm viva 1.1 (ProLab) User Manual

Page 520

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

508

■■■■■■■■

viva 1.1 (for process analysis)

Poll
The character string (optional) defined in Command is sent to the
instrument and an answer is expected and evaluated within the Time-
out
defined in the instrument settings. This action continues to be
repeated until a character string received corresponds to the regular
expression in the Comparison. If this does not take place within the
Timeout defined in the command, the command will be canceled.

Command

Character string that is to be sent to the instrument. Details are to be
found in the technical documentation of the instrument in question.

Entry

250 characters

NOTE

All ASCII characters can be entered with their two-digit hex code after
backslash (e.g. \1B = Escape).

Variable

Selection of an available sample data variable under which the received
data is to be saved for the actions Read, Wait for and Poll. Sample data
variables that are not permitted to be assigned to one of the specified
sample data variables must be defined for this purpose in the START
command (see Chapter 5.5.2.9.2.5, page 581).

Selection

Sample data variables

NOTE

Several indexed variables (e.g. Input1, Input2, etc.) can also be cre-
ated in order to be able to store the substrings requested with regular
expressions directly in various variables.

Example

Variable = SD.Input1

Comparison = (\d+\.\d+),(\d+\.\d+)

Two measured values (separated by a comma) are filtered out of a ran-
dom receiving line and saved in the two variables SD.Input1 and
SD.Input2.