4 cvs - sweep, Sweep – Metrohm viva 1.1 (ProLab) User Manual

Page 630

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

618

■■■■■■■■

viva 1.1 (for process analysis)

Potential 2

Potential that is applied to the electrodes during the Waiting time 2.
This potential can be used e.g. as deposition potential for the electro-
chemical deposition in stripping voltammetry.

Input range

–5.00 - 5.00 V (Increment: 0.001)

Selection

off

Default value

off

Waiting time 2

Waiting time during which the Potential 2 potential is applied to the
electrodes.

Only editable for Potential 2

off, otherwise inactive.

Input range

0.0 - 9999.9 s (Increment: 0.1)

Default value

0.0 s

Potential 3

Potential that is applied to the electrodes during the Waiting time 3.

Input range

–5.00 - 5.00 V (Increment: 0.001)

Selection

off

Default value

off

Waiting time 3

Waiting time during which the Potential 3 potential is applied to the
electrodes.

Only editable for Potential 3

off, otherwise inactive.

Input range

0.0 - 9999.9 s (Increment: 0.1)

Default value

0.0 s

Equilibration time

Waiting time before the sweep during which the Start potential is
applied to the electrodes.

Input range

0.0 - 9999.9 s (Increment: 0.1)

Default value

5.0 s

5.5.2.11.2.4

CVS - Sweep

Tab: Method

CVS Properties... Sweep

Command name

Name of the command.

Entry

25 characters