4 properties - special beaker, Special beakers – Metrohm viva 1.1 (ProLab) User Manual

Page 848

Advertising
background image

6.12 Rack data subwindow

■■■■■■■■■■■■■■■■■■■■■■

836

■■■■■■■■

viva 1.1 (for process analysis)

Input range

0 - 235 mm

Tower 2
Defines the lift positions for tower 2. These apply for all rack positions
except those that are defined as Special beakers.

Work position

Work position for lift 2.

Input range

0 - 235 mm

Rinse position

Rinse position for lift 2.

Input range

0 - 235 mm

Shift position

Shift position for Lift 2. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then it will shift at the current
lift position. This means that the shift position must be selected so that a
safe movement across the entire rack is possible at any time.

Input range

0 - 235 mm

Special position

Special position for lift 2.

Input range

0 - 235 mm

6.12.3.4

Properties - Special beaker

Dialog window: Configuration

Rack data [Edit] Properties... Rack

data

Special beakers

The table shows the data of all special beakers of the attached rack in tab-
ular form. Any position on the rack can be assigned to each special
beaker. The table cannot be edited directly.

NOTE

Special beakers should preferably be set at high rack positions in order
to be able to begin sample series at rack position 1. Rack positions
defined as special beakers are no longer available as sample positions. A
specific work position can be defined for each special beaker at tower 1
and tower 2 (if present). The shift, rinse and special positions of the cor-
responding tower are applied from the general rack positions.