4 cpvs - sweep, Sweep – Metrohm viva 1.1 (ProLab) User Manual

Page 648

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

636

■■■■■■■■

viva 1.1 (for process analysis)

5.5.2.11.3.4

CPVS - Sweep

Tab: Method

CPVS Properties... Sweep

Command name

Name of the command.

Entry

25 characters

Equilibration potential

Potential applied during the Equilibration time.

Input range

–5.000 - 5.000 V (Increment: 0.001)

Default value

0.450 V

Equilibration time

Time during which the Equilibration potential is applied to the elec-
trodes.

Input range

0.0 - 50.0 s (Increment: 0.1)

Default value

5.0 s

Plating potential

Potential applied during the Plating time.

Input range

–5.000 - 5.000 V (Increment: 0.001)

Default value

–0.175 V

Plating time

Time during which the Plating potential is applied to the electrodes.

Input range

0.0 - 50.0 s (Increment: 0.1)

Default value

4.0 s

Stripping
In this table, you can define up to five stripping potentials, which are suc-
cessively applied during the corresponding stripping time. By default, three
stripping potentials are specified.

Stripping potential

Potential applied during the Stripping time.

Stripping time

Time during which the Stripping potential is applied to the electrodes.