2 add std - properties – Metrohm viva 1.1 (ProLab) User Manual

Page 465

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

453

5.5.2.4.3.2

ADD STD - Properties

Dialog window: Method

ADD STD Properties... ADD STD - 'Command name'

Command name

Name of the command.

Entry

25 characters

Standard

Solution

Entry of a solution name or selection of an auxiliary solution from the
method or from the solution table. The selection of a solution from the
solution table is necessary only if the solution is to be added automatically
with a dosing device or if the working life of the solution is to be checked
at the start of the method.

Entry

1 - 25 characters

Selection

'Standard name' | 'empty'

Default value

'empty'

'Standard name'
The standard solutions defined in the method appear first for selection,
followed by the standard solutions present in the solution table (Con-
figuration) (each time in alphabetical order).

Addition increments

Number

Number of variable volume increments that are to be added. The first vol-
ume is added the first time the command is called, the second volume at
the second time, etc. If the command is called more often than volumes
were defined, then the last volume is added in each case.

Input range

1 - 25 (Increment: 1)

Default value

1

Addition volume #

Size of the individual volume that is to be dosed in. The number of addi-
tion volumes is defined by the amount defined in Number. If the com-
mand is called more often than volumes were defined, then the last vol-
ume is added in each case.

Input range

0.010 - 1,000.000 mL (Increment: 0.001)

Default value

0.100 mL

Addition
Selection of the addition mode for dosing standard solutions.

Advertising