Metrohm viva 1.1 (ProLab) User Manual
Page 621
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.1 (for process analysis)
■■■■■■■■
609
Post-treatment
1 Cleaning potential
Cleaning potential is applied to the electrodes.
2 Cleaning time
Idle until the period defined in the Cleaning time parameter has
expired.
only for hydrody-
namic measurement
= on
3 Stirrer off
The stirrer is switched off.
only for standby
potential = off
4 Potential off
The potential applied to the electrodes is switched off.
only for standby
potential
≠
off
5 Standby potential on
Standby potential is applied to the electrodes.
Parameters
The parameters for the CVS command are set on the following tabs:
■
■
■
■
■
Command variables
The following command variables are generated by the CVS command in
the method run and can be used in formulas under the designation 'Com-
mand name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
DBL
Total duration for the processing of the command in
s