8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

962

■■■■■■■■

viva 1.1 (for process analysis)

NOTE

The dosing rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Input range

0.01 - 166.00 mL/min

Default value

4.00 mL/min

Filling rate

The dosing cylinder will be filled at this rate after dosing. The maximum
filling rate depends on the cylinder volume of the dosing unit used. If the
entered filling rate is too high for the selected dosing device, it will auto-
matically be reduced during filling to the largest possible value.

NOTE

The filling rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Input range

0.01 - 166.00 mL/min

Default value

4.00 mL/min

Fill automatically

on | off (Default value: on)

If this option is activated, the dosing cylinder is filled automatically after
dosing. The volume display shows the added volume until the filling pro-
cedure has ended. If this option is disabled, the added volume is displayed
cumulatively.

[Start]

Starts fixed volume dosing for the selected dosing device. The dosed vol-
ume is displayed live.

NOTE

Parameters that are modified after dosing has started are not valid until
the next dosing procedure.