5 Method

viva 1.1 (for process analysis)

■■■■■■■■

429

Move angle

This parameter specifies the angle by which the sample rack is to be rota-
ted relative to its momentary position. It is visible only when Move = Rel-
ative angle
.

814, 815, 858
Input range

-359.90 - 359.90 °

Default value

5.00 °

Beaker test

Selection of the action to be carried out when the beaker sensor defined
for the rack detects no beaker at the target position. If the robotic arm is
selected as beaker sensor, then a Swing Head must be installed. A work
position that is also defined in the rack data must be defined in such a
way that a contact exists between beaker sensor and beaker. This work
position will be automatically moved to for beaker recognition after the
command MOVE. This section is not visible when Move = Relative
angle
.

Selection

Display message | Cancel determination | Can-
cel determination and series

Default value

Display message

Display message
A message is displayed, the determination is canceled and the next
determination of the series is started after confirmation of the mes-
sage. If an exit track is present, it will be called. No error track will be
called.
Cancel determination
The determination is canceled and the next determination of the series
is started. If an exit track is present, it will be called. No error track will
be called.
Cancel determination and series
Both the determination and the series are stopped. If an exit track is
present, it will not be called. An error track will be called.

Parameters

Shift rate

Selection of the rate at which the rack rotates.

Input range

3 - 20 °/s

Default value

20 °/s