5 Method

viva 1.1 (for process analysis)

■■■■■■■■

649

Standby potential

Standby potential

The standby potential is applied to the electrodes after completion of the
sweep. It remains in effect until it is switched off manually or until a new
potential is applied to the electrodes. If no standby potential is defined,
then the potential applied to the electrodes is automatically switched off
after completion of the sweep.

Input range

–5.000 - 5.000 V (Increment: 0.001)

Selection

off

Default value

off

5.5.2.11.4.6

CP - Galvanostat

Tab: Method

CP Properties... Galvanostat

Command name

Name of the command.

Entry

25 characters

Potential measuring range

Maximum range

Limitation of the highest potential measuring range.

Selection

5 V | 5 mV | 50 mV | 500 mV

Default value

5 V

Minimum range

Limitation of the lowest potential measuring range.

Selection

5 mV | 50 mV | 500 mV | 5 V

Default value

50 mV