7 Devices

viva 1.1 (for process analysis)

■■■■■■■■

911

Assigned ports

Selection of an output port assigned to an input port. The signal of the
output port depends on the signal of the input port.

Only active and editable for the port types Level detector and Leak
detector.
Selection

All digital outputs

NOTE

Digital outputs of the Type = Alarm signal cannot be selected.

Signal

Selection of the value which the signal on the output port should have.

Only active and editable for the port types Level detector and Leak
detector.
Selection

Input signal | Invertet input signal

Default value

Input signal

Signal processing

Delayed edge

The input signal will only be recorded with a delay time on the rising or
falling edge.

Selection

ascending | descending

Default value

descending

Delay time

Time that has to elapse before the input signal on the rising or falling
edge is recorded.

Input range

0 - 214,783,647

Default value

0

Invert signal

on | off (Default value: off)

If this check box is activated, then the input signal received will be inver-
ted.

Message
Only active and editable for Type = Level detector.