Metrohm viva 1.1 (ProLab) User Manual
Page 950
7.10 IO controller
■■■■■■■■■■■■■■■■■■■■■■
938
■■■■■■■■
viva 1.1 (for process analysis)
Working range
for Control task = Digestion unit
Input range
0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'
Default value
40.0 'Mapping unit of the input or of the cal-
culated value'
for Control task = Free controller
Input range
0 - Smallest distance between setpoint value
and 0% mapping or 100% mapping 'Map-
ping unit of the input or of the calculated
value'
Default value
10 'Mapping unit of the input or of the calcu-
lated value'
Output parameters
PWM period
for Control task = Digestion unit
Input range
0 - 9,999.0 s
Default value
1.0 s
for Control task = Free controller
Input range
0 - 9,999.0 s
Default value
10 s
Min. pulse ratio
for Control task = Digestion unit
Input range
0 - 100 %
Default value
0 %
for Control task = Free controller
Input range
0 - 100 %
Default value
10 %
max. pulse ratio
for Control task = Digestion unit
Input range
0 - 100 %
Default value
35 %