8 Manual control

viva 1.1 (for process analysis)

■■■■■■■■

981

Start delay

Waiting time in seconds or minutes before the initiated action is started.

Input range

0 - 999999

Default value

0

Selection

s | min

Default value

s

Working time

Time in seconds or minutes during which the digital output is set to
ACTIVE. After the working time has elapsed, it is set to INACTIVE again.

Input range

0 - 999,999

Default value

0.01

Selection

s | min

Default value

s

End delay

Waiting time after the working time has expired

Input range

0 - 999999

Default value

0

Selection

s | min

Default value

s

Status

Display of the current status.

Selection

ACTIVE | INACTIVE | ERROR

Time to finish

Shows the remaining running time of the timer control.

[Start]

Sets the digital output to ACTIVE.

[Stop]

Sets the digital output to INACTIVE.

Set default

Port description

Display of the port description.

Port

Display of the internal description of the port.