Metrohm viva 1.1 (ProLab) User Manual

Page 649

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1 (for process analysis)

■■■■■■■■

637

[New]

Opens the Stripping dialog window, in which the stripping potential and
the stripping time of a potential step can be entered ((see Chapter
5.5.2.11.3.7, page 639)
).

[Properties]

Opens the Stripping dialog window, in which the stripping potential and
the stripping time of a potential step can be edited (see Chapter
5.5.2.11.3.7, page 639)
.

[Delete]

The selected potential step is deleted.

Cleaning potential

Potential that is applied to the electrodes during the Cleaning time.

Input range

–5.000 - 5.000 V (Increment: 0.001)

Default value

1.625 V

Cleaning time

Waiting time during which the Cleaning potential is applied to the elec-
trodes.

Input range

0.0 - 50.0 s (Increment: 0.1)

Default value

5.0 s

Measuring interval

Measuring interval for data acquisition during plating and stripping.

Selection

1.28 ms | 2.56 ms | 5.12 ms | 10.24 ms | 20.48
ms
| 40.96 ms | 81.92 ms

Default value

5.12 ms

Preparation cycles

Number of cycles without current measurement that are run through
before the measuring cycles.

Input range

0 - 200 (Increment: 1)

Default value

2

Measuring cycles

Number of cycles with current measurement.

Input range

1 - 50

Default value

2

Sweep duration

Shows the duration of the sweep in s.