Waveform groups – Fluke Biomedical medTester 5000C User Manual

Page 110

Advertising
background image

medTester 5000C
Operators Manual

7-2

Waveform Groups

You can generate any of the waveforms listed in the five groups in the table
below for the testing of patient monitors. The table’s first column lists the
names of the waveforms as they appear in the medTester 5000C display.
Column two describes the display names, and column three shows the remote
command for each waveform.

WAVEFORM GROUPS

medTester 5000C

Display Wave Menu

Name

Description

Remote Command

Group Name:

PERF

2Hz SQ

2 Hz Square

W001

4 Sec

4 Second Pulse

W002

10Hz

10 Hz Sine Wave

W003

40Hz

40 Hz Sine Wave

W004

60Hz

60 Hz Sine Wave

W005

100Hz

100 Hz Sine Wave

W006

1KHz

1 kHz Square Wave

W007

1HzTRI

2 Hz Triangle

W008

CMRR

60 Hz Common Mode Rejection Test

W009

.5Hz

0.5 Hz Sine Wave

W014

Group Name: ECG

30BPM

Normal Sinus ECG at 30 Beats Per
Minute (BPM)

W010

60BPM

Normal Sinus ECG at 60 BPM

W011

120BPM

Normal Sinus ECG at 120 BPM

W012

240BPM

Normal Sinus ECG at 240 BPM

W013

Group Name: ARRHY1

AFIB

Atrial Fibrillation 1, Course

W015

2°BLK1

2nd Degree A-V Block, Type 1,
Wenckebach

W016

PAC

Premature Atrial Contractions

W017

MB80

Missed Beat at 80 BPM

W018

MB120

Missed Beat at 120 BPM

W019

Group Name: ARRHY2

PVC1

PVC Type 1, Left

W020

PVC2

PVC Type 2, Right

W021

MFPVC Multi-focal

PVCs

W022

PVCRT

PVC 1, R on T

W023