Running Waveforms from The WAVES Menu

To turn on any of the waveforms, follow these instructions:

1.

Attach the patient leads from the EUT (patient monitor) to
the binding posts on the medTester 5000C.

2.

Power on the EUT. Note that, for your convenience, you can
plug the patient monitor into the medTester’s test receptacle
since test receptacle power is on when running any waveform.

3.

Press the desired medTester function keys to access the wave
menu item. The medTester outputs the selected waveform to
the patient monitor.

Note

You can change waves in a group by selecting a new wave, or
you can press

Esc

and change to a wave in another group. Waves

continue to run when changing groups. Press

Esc

twice to exit

waves and turn off the running waveform.