Posttest prompts, Customizing spo2 autosequences, Make your own autosequence – Fluke Biomedical medTester 5000C User Manual

Page 246: For cardiosat 100

Advertising
background image

medTester 5000C
Operators Manual

15-18

For CardioSat 100:

SPO2 Autosequence Number: A74

Autosequence Name: GENERC

Manufacturer: Generic–Not Specified

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

1 99

± 4

250 100

2 96

± 4

230 100

3 92

± 4

205 100

4 88

± 5

180 100

5 78

± 5

155 100

6 75

± 5

130 100

7 65

± 6

105 100

8 55

± 6

80 100

9 45

± 6

55 100

10 35

± 6

30 100

11 90

± 4

95 100

12 88

± 5

45 100

13 85

± 5

130 20

14 92

± 4

95 40

15 93

± 4

90 50

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test