Posttest prompts, Customizing spo2 autosequences, Make your own autosequence – Fluke Biomedical medTester 5000C User Manual
Page 246: For cardiosat 100
![background image](/files/785336/content/doc246.png)
medTester 5000C
Operators Manual
15-18
For CardioSat 100:
SPO2 Autosequence Number: A74
Autosequence Name: GENERC
Manufacturer: Generic–Not Specified
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent)
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
1 99
± 4
250 100
2 96
± 4
230 100
3 92
± 4
205 100
4 88
± 5
180 100
5 78
± 5
155 100
6 75
± 5
130 100
7 65
± 6
105 100
8 55
± 6
80 100
9 45
± 6
55 100
10 35
± 6
30 100
11 90
± 4
95 100
12 88
± 5
45 100
13 85
± 5
130 20
14 92
± 4
95 40
15 93
± 4
90 50
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test