Biomedical Test System

Transcutaneous Pacemaker (Pacer) Module

16

16-13

Pacer Autosequence Number: A23

Autosequence Name: MDE3PM

Load: Adapter Module TQA-6, External 600 Ohms (Note: In Step 11 substitute 1000

Ohms.)

Step Number

Current

(in milliamperes)

Rate

(in PPM, pulses per minute)

1 60

40

2 60

60

3 60

70

4 60

75

5 60

80

6 60

90

7 60

100

8 60

120

9 60

140

10 60

160

11 120

60

12 90

100

13 120

100

Pulse Tests:

Current Limit: ± 20%

Rate Limit: ± 5 PPM

Demand Test: 80 PPM

Underdrive: 30 BPM

Ovedrive: 160 BPM

Async Test: 80 PPM

Overdrive: 160 BPM

Refractory Period Test: No