Reinitialize factory default settings – Fluke Biomedical medTester 5000C User Manual
Page 269
Advertising
![background image](/files/785336/content/doc269.png)
Biomedical Test System
Transcutaneous Pacemaker (Pacer) Module
16
16-13
Pacer Autosequence Number: A23
Autosequence Name: MDE3PM
Load: Adapter Module TQA-6, External 600 Ohms (Note: In Step 11 substitute 1000
Ohms.)
Step Number
Current
(in milliamperes)
Rate
(in PPM, pulses per minute)
1 60
40
2 60
60
3 60
70
4 60
75
5 60
80
6 60
90
7 60
100
8 60
120
9 60
140
10 60
160
11 120
60
12 90
100
13 120
100
Pulse Tests:
• Current Limit: ± 20%
• Rate Limit: ± 5 PPM
Demand Test: 80 PPM
• Underdrive: 30 BPM
• Ovedrive: 160 BPM
Async Test: 80 PPM
• Overdrive: 160 BPM
Refractory Period Test: No