Fluke Biomedical medTester 5000C User Manual
Page 208
medTester 5000C
Operators Manual
13-12
Note
Autosequences always communicate at a fixed baud rate;
therefore, you don't need to set the rate for any port used only
for autosequences.
To set a Wedge port baud rate:
1.
From the medTester 5000C main menu,
MENU 1
, press:
UTIL
F5
2.
From the utility menu, press:
BAUD
F1
to see this display:
Pressing the medTester’s right arrow lets you access the remaining port:
3.
Press the function key that corresponds to the port you want, for
example COM2-1.
4.
Press
F4
to change the baud rate value that appears in the display
until you reach the value you want:
COM PORT BAUD RATES
COM2-4
F5
COM2-3
F4
COM1
F1
COM2-2
F3
COM2-1
F2
COM PORT BAUD RATES
COM2-8
F4
COM2-5
F1
COM2-7
F3
COM2-6
F2
SET COM2-1 BAUD RATE:
9600
BAUD
F4
STORE
F5