For the index 2xl – Fluke Biomedical medTester 5000C User Manual

Page 248

Advertising
background image

medTester 5000C
Operators Manual

15-20

For the Index 2XL:

SPO2 Autosequence Number: A72

Autosequence Name: PMS

Manufacturer: PMS M1190

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Base TLC

1 96

± 3

75 6.0 3500

2 98

± 3

230 6.0 3500

3 92

± 3

75 6.0 3500

4 82

± 4

75 6.0 3500

5 72

± 5

75 6.0 3500

6 98

± 3

45 6.0 3500

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 6.0%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test

SPO2 Autosequence Number: A73

Autosequence Name: DATEX

Manufacturer: Datex

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Base TLC

1 96

± 3

75 6.0 3030

2 98

± 3

230 6.0 3030

3 92

± 3

75 6.0 3030

4 82

± 4

75 6.0 3030

5 72

± 5

75 6.0 3030

6 98

± 3

45 6.0 3030

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 6.0%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test