For a pca test, Ipt-1 – Fluke Biomedical medTester 5000C User Manual

Page 158

Advertising
background image

medTester 5000C
Operators Manual

10-20

IPT-1 AUTOSEQUENCE TESTS

Nominal

Flow Rate

(ml per

hour)

Test Type

High
Limit

(ml per

hour)

Low

Limit

(ml per

hour)

End Type

(manual,

volume,

or timed)

Ending Quantity

If Applicable

(ml for volume or

hours:minutes

for timed)

Autosequence Name (Test Type): LINEAR

Keyboard Shortcut Command: A50

60.0 Steady

Flow

66.0 54.0 Volume

5.0

120.0 Steady

Flow

132.0 108.0 Volume

10.0

240 Steady

Flow

264.0 216.0 Volume

20.0

Autosequence Name (Test Type): PULSED

Keyboard Shortcut Command: A51

60.0 High

Accuracy

66.0 54.0 Volume

5.0

120.0 High

Accuracy

132.0 108.0 Volume

10.0

120.0 High

Accuracy

132.0 108.0 Volume

10.0

Autosequence Name (Test Type): LOW

Keyboard Shortcut Command: A52

10.0 Steady

Flow

11.0 9.0 Timed

00:05

Autosequence Name (Test Type): HIGH

Keyboard Shortcut Command: A53

480.0 High

Accuracy

504.0 456.0 Timed

00:05

Autosequence Name (Test Type): EVAL

Keyboard Shortcut Command: A54

5.0 Steady

Flow

5.9 4.1 Volume

1.5

10.0 Steady

Flow

11.0 9.0 Volume

2.5

20.0 High

Accuracy

22.0 18.0 Volume

2.5

30.0 High

Accuracy

33.0 27.0 Volume

2.5

60.0 High

Accuracy

66.0 54.0 Volume

5.0

120.0 High

Accuracy

132.0 108.0 Volume

10.0

240.0 High

Accuracy

264.0 216.0 Volume

30.0

480.0 High

Accuracy

528.0 432.0 Volume

60.0