For cardiosat 100 – Fluke Biomedical medTester 5000C User Manual

Page 244

Advertising
background image

medTester 5000C
Operators Manual

15-16

For CardioSat 100:

SPO2 Autosequence Number: A70

Autosequence Name: NELLCR

Manufacturer: Nellcor

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

1 96

± 3

75 100

2 98

± 3

230 100

3 81

± 4

38 25

4 81

± 4

112 100

5 81

± 4

201 100

6 92

± 3

75 100

7 82

± 4

75 100

8 72

± 5

75 100

9 98

± 3

30 100

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test

SPO2 Autosequence Number: A71

Autosequence Name: OHMEDA

Manufacturer: Ohmeda 3700

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

1 96

± 3

75 100

2 98

± 3

230 100

3 92

± 3

75 100

4 82

± 4

75 100

5 72

± 5

75 100

6 98

± 3

45 100

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test

Advertising