Print a range of records, Print all records, Print records by type – Fluke Biomedical medTester 5000C User Manual

Page 113: Conducting waveform tests in safety autosequences

Advertising
background image

Biomedical Test System

Performance Waves

7

7-5

Conducting Waveform Tests in Safety
Autosequences

You can execute performance wave tests at the end of any safety
autosequence. The medTester 5000C prompts you to begin or to skip these
tests.

When performance waveforms are generated within a safety autosequence,
they follow this sequence for the listed periods of time:

Performance Waveforms in Safety Autosequences

Waveform Name

Test Duration (In Seconds)

2 Hz Square

Continuous

4 sec Step

8

10 Hz Sine

4

40 Hz Sine

4

60 Hz Sine

60

100 Hz Sine

4

1 kHz Square

4

2 Hz Triangle

Continuous

CMRR 60 Hz

Continuous

ECG 60 BPM

Continuous

ECG 30 BPM

30

ECG 60 BPM

30

ECG 120 BPM

30

ECG 240 BPM

30

Note

Tests with a predefined duration automatically end or can be
terminated at any time by pressing

(NEXT)

. Those marked

continuous can be terminated by pressing

NEXT

.