Biomedical Test System

Autosequences

5

5-19

End-of-Test Prompts

At the end of the autosequence, medTester prompts you for the following
data: (prompts can be turned off)

COMMENTS

NEXT TEST DUE DATE

USER TIME

.

STORE RECORD?

(if memory was selected), or

PRINT?

(if print was

selected

Toolbox Monitors

The toolbox menu on the medTester 5000C main menu contains monitoring
sequences for temperature and relative humidity. To add these two options to
your medTester, contact Extech Instruments Coporation.

Temperature Monitor
You can monitor temperature readings over periods of time, print the results,
and store the results in memory. To access the monitor from the keyboard,
enter:

TEMPMON

F4

TOOLBOX

F2

Selections then display as:

Select

MEMORY

if you want to store the test record. This requires both the RS-

232/Printer module and the 100 Record Memory module and . Select

PRINT

if

you want to print the test record. Select

BOTH

if you want to store and print

the record. Follow these set-up instructions:

1.

Plug the temperature adapter test leads into the

EXT INPUT

jacks

on medTester’s front panel.

2.

Set the adapter’s power switch to the desired temperature scale,
Fahrenheit or Celsius. It must match the medTester selection.

3.

Press the

°F/°C (F2)

key for the desired temperature scale.


RUN TEMPERATURE MONITOR A18 TO:

MEMORY

F1

PRINT

F2

NEITHER

F4

BOTH

F3