External resistance, Isolated power and ground fault test – Fluke Biomedical medTester 5000C User Manual

Page 52

Advertising
background image

medTester 5000C
Operators Manual

4-6

RL

—right leg,

RA

—right arm,

LA

—left arm,

LL

—left leg, and

V

—V1 through V6 leads tied together. You can measure V leads

individually.
Displayed measurements are in

µA.

Note

If you receive abnormally high leakage current measurements
with V1-V6 leads tied together, you can inspect individual V
leads by removing the other leads. This procedure allows you to
isolate the lead with a high leakage current reading.

Lead to Lead Leakage

See Figure 4-6, ECG Interlead Leakage.

1.

Plug the EUT into the

TEST RECEPTACLE.

2.

Connect desired ECG leads to their respective binding posts on
the medTester 5000C.

3.

Enter menu commands:

MANUAL

F2

µA

F2

ECG

F4

LD/LD

F3

4.

Select the desired lead.

This measures leakage between a selected lead and all other leads tied
together.