Fluke Biomedical medTester 5000C User Manual

Page 238

Advertising
background image

medTester 5000C
Operators Manual

15-10

Each autosequence can contain up to ten SPO2 simulation steps, except the
INDEX 2XL which can contain up to 15 SPO2 steps. After those, several
special tests will be performed if they have been programmed into the
autosequence.

SPO2 Simulation Steps

For each step the medTester commands the SPO2 simulator to set up a
simulation of the programmed SPO2 in percentage, heart rate in beats per
minute (BPM), and amplitude in percentage. For a given step, only the
commands necessary for a change are sent.

For the Oxitest Plus, the medTester also sets up the programmed artifact.

For the CardioSat 100 and Index 2XL, the commands sent to the SPO2
simulator are noted on the medTester display to show you what is happening.

For the Index 2XL, the medTester can also be set up for Motion-Preset Tests
based on either Level 1 or Level 2 tests. The tables below show the eight
motion presets for each of the two levels of tests.

L

EVEL

1

P

RESET

N

AME

S

P

O2

% R

ATE

BPM

08 Normal/Tap

98

55

09 Normal/Shiver

98

55

10 Weak/Tap

90

95

11 Weak/Shiver

90

94

12 Brad/Tap

88

45

13 Hypoxic/Tap

70

95

14 Hypoxic/Shiver

70

95

15 Neonate/Shiver

90

180

L

EVEL

2

P

RESET

N

AME

S

P

O2

% R

ATE

BPM

16 BradyTap#2

88

45

17 Hypox/Tap#2

70

95

18 WeakTap#2

80

95

19 NormalTap#2

93

55

20 Asystole

91

90

21 LowFreq1

80

75

22 LowFreq2

70

75

23 SlowTap

80

75


For the CardioSat 100 and the INDEX 2XL, the commands sent to the
SPO2 simulator are noted on the medTester display to show you what is
happening.

The medTester waits for the programmed number of heartbeats for a settling
time for the pulse oximeter. The medTester beeps when settling is finished.