Fluke Biomedical medTester 5000C User Manual

Page 26

Advertising
background image

medTester 5000C
Operators Manual

1-14

Toolbox

Toolbox allows for the use of external measurement adapters for:

Tachometer—with a range of 100 to 20,000 RPM.

Temperature—with a range of 0 to 200

°F or

0 to 100

°C.

Humidity—ranging from 15% to 90% relative.

For more information about Toolbox, see Chapter 4, Manual Tests.

Test Receptacle
The medTester 5000C has a test receptacle on the top panel into which the
equipment under test (EUT) is plugged. The test receptacle supplies power to
the EUT. The EUT must be powered through the test receptacle whenever
you measure the following:

Leakage current—All case and ECG lead measurements.

Equipment current.

Power cord resistance.

Test receptacle power is normal except during case and ECG leakage tests
(not including isolation tests). For these tests you can set the power to the
following conditions, each of which is indicated by annunciator LEDs on the
top panel:

Ground—closed or open.

Polarity—normal or reversed.

Neutral—closed or open.

Note

During case internal leakage, ground is always open.

Current is supplied at a maximum instantaneous level of 20 amps.

CSA Label

On the back of the medTester 5000C, you can see a label which specifies the
conditions under which the test receptacle can operate:

1840 VA, 16 A continuous operation.

Up to 19 A for no longer than two (2) minutes. Then powered off
for eight (8) minutes (20 percent of duty cycle).