Biomedical Test System

SPO2 Module

15

15-17

For CardioSat 100:

SPO2 Autosequence Number: A72

Autosequence Name: HPMRLN

Manufacturer: Hewlett Packard Merlin Series

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

1 96

± 3

75 100

2 98

± 3

230 100

3 92

± 3

75 100

4 82

± 4

75 100

5 72

± 5

75 100

6 98

± 3

45 100

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test

SPO2 Autosequence Number: A73

Autosequence Name: DATEX

Manufacturer: Datex

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

1 96

± 3

75 100

2 98

± 3

230 100

3 92

± 3

75 100

4 82

± 4

75 100

5 72

± 5

75 100

6 98

± 3

45 100

Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test