Hot to neutral, Neutral to ground, Hot to ground – Fluke Biomedical medTester 5000C User Manual

Page 44: Leakage current

Advertising
background image

medTester 5000C
Operators Manual

3-4

16.

COM1 PORT

—Receives input locally from the optional keyboard,

bar code scanning gun, or personal computer or terminal. Outputs
data to a personal computer or terminal. With a Wedge adapter,
this port is used by the external PC keyboard interface.

17.

FUSE

—Twenty (20) ampere, slow-blow type, 250 V.

18.

POWER SWITCH

—Powers the medTester 5000C on or off.

19.

POWER CORD and PLUG

The hardwired power plug is a 20A

configuration. For most testing leave the 20A to 15A adapter plug
on to plug into a 15A outlet.

Power Up
Locate the power switch on the rear panel, and power on the
medTester 5000C. The name and revision appear followed by the main menu.
To darken or lighten the display, use the

DISPLAY

keys to the left of the

display. The black-circled key darkens the display. The white-circled key
lightens the display.

Navigating the Menus
Operating the medTester 5000C is as easy as pressing the function key on the
top panel that corresponds to the desired menu item in the display. There are
five function keys marked

F1

through

F5

below the display (see the figure

below). The function keys select the menu functions that appear in the display
just above them. The medTester 5000C is operated through menu choices in
a tree-like structure. When you power on the medTester, the first menu you
see sits at the top of the tree. As you make menu choices, you move to lower
levels, or branches, of the tree. The menu you see below is the main menu.

Note the right arrow located at the right edge of the menu. Whenever you see
a right arrow on a menu, it indicates that there are more menu items at the
same level. Access those items by pressing the right arrow key on the
keyboard. What you then see on the display is:

00/00/00

MENU

1

00:00:00

MANUAL

F2

AUTO

F1

WAVES

F3

CHECK

F4

UTIL

F5