Downloading – Fluke Biomedical medTester 5000C User Manual

Page 185

Advertising
background image

Biomedical Test System

Remote Operation

12

12-7

Forcing A Port

An alternative to the above procedure is forcing a port. For example, if you
want COM1 to be the remote port, type or send

remote1

. This will force

COM1 to act as the remote port. To force COM2, type or send

remote2

.

If you have made a remote connection, the medTester responds by sending:

REMOTE MODE

out the remote serial port. At the same time, the

medTester 5000C display becomes blank.

Command Syntax Rules

•

Send commands in uppercase or lowercase letters.

•

Spaces are ignored.

•

Use the backspace character for editing.

•

Terminate commands with a carriage return (CR), or line feed (LF),
or both.

•

Commands with an equal sign (=) require that at least one parameter
be included with the command.

•

Use commas to separate multiple parameters.

Complete lists of medTester 5000C remote commands and parameters are
shown in the Remote Commands List section later in this chapter.

Responses to Commands

•

Every command you issue to the medTester returns a response to
the remote serial port.

•

Responses return character strings, numbers, or data files.

•

If no other response appears, an asterisk (*) appears to indicate that
the command has been executed.

•

Previous commands must execute and return their responses before
the medTester will accept further commands.

•

The null command (consisting only of terminating characters)
returns a question mark (?).

•

Command responses are in uppercase letters followed by a carriage
return and line feed.

•

Measurement reading responses appear as they do on the
medTester 5000C. Numbers may appear with or without a decimal
point and are followed by the unit of measurement abbreviation.
See the table that follows.