Biomedical Test System

Remote Operation

12

12-7

Forcing A Port

An alternative to the above procedure is forcing a port. For example, if you
want COM1 to be the remote port, type or send

remote1

. This will force

COM1 to act as the remote port. To force COM2, type or send

remote2

.

If you have made a remote connection, the medTester responds by sending:

REMOTE MODE

out the remote serial port. At the same time, the

medTester 5000C display becomes blank.

Command Syntax Rules

Send commands in uppercase or lowercase letters.

Spaces are ignored.

Use the backspace character for editing.

Terminate commands with a carriage return (CR), or line feed (LF),
or both.

Commands with an equal sign (=) require that at least one parameter
be included with the command.

Use commas to separate multiple parameters.

Complete lists of medTester 5000C remote commands and parameters are
shown in the Remote Commands List section later in this chapter.

Responses to Commands

Every command you issue to the medTester returns a response to
the remote serial port.

Responses return character strings, numbers, or data files.

If no other response appears, an asterisk (*) appears to indicate that
the command has been executed.

Previous commands must execute and return their responses before
the medTester will accept further commands.

The null command (consisting only of terminating characters)
returns a question mark (?).

Command responses are in uppercase letters followed by a carriage
return and line feed.

Measurement reading responses appear as they do on the
medTester 5000C. Numbers may appear with or without a decimal
point and are followed by the unit of measurement abbreviation.
See the table that follows.