Biomedical Test System
Transcutaneous Pacemaker (Pacer) Module
16
16-17
Reinitialize Factory Default Settings
Any customized autosequences can be returned to the factory default settings.
To do this, enter these medTester commands:
1.
From the medTester 5000C main menu,
MENU
1, press the
right arrow to access
MENU2
.
2.
Enter the command:
CUSTOM
F3
3.
From the custom menu, press the right arrow and enter the
command:
INIT
F3
4.
Press the right arrow, and then press:
PACER
F1
5.
To initialize pacer autosequences to defaults, press:
YES
F5