Biomedical Test System

Manual Tests

4

4-9

Power Cord Resistance

See Figure 4-9, Power Cord Resistance Test.

1.

Plug the equipment under test into the

TEST RECEPTACLE

at the

top of the medTester 5000C.

2.

Connect the dual banana-plug end of a Kelvin cable to the two
red external meter posts.

3.

Connect the alligator-clip end of the Kelvin cable to a grounded
point on the case of the equipment under test.

4.

Enter menu commands:

OHMS

F4

MANUAL

F2

PWRCORD

F1

The

CURRENT SOURCE

lamp lights up to indicate that the

connection is made. The display reads the resistance (in OHMS)
from the case to the power cord ground pin.