Biomedical Test System
SPO2 Module
15
15-21
For the Index 2XL:
SPO2 Autosequence Number: A74
Autosequence Name: GENERC
Manufacturer: Generic–Not Specified
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Base TLC
1 99
± 4
250 6.0 4000
2 96
± 4
230 6.0 4000
3 92
± 4
205 6.0 4000
4 88
± 5
180 6.0 4000
5 78
± 5
155 6.0 4000
6 75
± 5
130 6.0 4000
7 65
± 6
105 6.0 4000
8 55
± 6
80 6.0 4000
9 45
± 6
55 6.0 4000
10 35
± 6
30 6.0 4000
11 90
± 4
95 6.0 4000
12 88
± 5
45 6.0 4000
13 85
± 5
130 2.0 4000
14 92
± 4
95 3.0 4000
15 93
± 4
90 4.0 4000
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 6.0%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
No probe test included