Biomedical Test System

Manual Tests

4

4-7

Isolation Lead Leakage

See Figure 4-7, ECG Lead Isolation Test: All Leads and Individual Leads.

You can do the following two tests in an all-lead or individual-lead mode. The
selected lead or leads are connected to the AAMI load with line voltage
applied to the lead(s). The test receptacle is fixed with normal polarity, closed
ground, and closed neutral during these tests.

WARNING!

During any isolation lead leakage test, 120 V at 1 mA is present
on the ECG leads. This is an electrical shock hazard.

Lead Isolation—All Leads

1.

Plug the EUT into the

TEST RECEPTACLE.

2.

Connect the ECG leads to be tested to their respective, color-
coded binding posts on the medTester 5000C.

3.

Enter menu commands:

µA mV

F2

MANUAL

F2

ECG

F4

ALL/ISO

F4

3.

To begin the measurement, press

ISOV ON

F5

. The test begins,

and the

ISO VOLTS

annunciator lights. The reading is displayed for

up to 15 seconds. You can escape to terminate the test. The test
self-terminates after 15 seconds.

Lead Isolation—Single Lead

1.

Plug the EUT into the

TEST RECEPTACLE.

2.

Connect the ECG lead to be tested to its respective color-coded
binding post on the medTester 5000C.

3.

Enter menu commands:

µA mV

F2

MANUAL

F2

ECG

F4

LD/ISO

F5