Biomedical Test System

The Wedge Adapter

13

13-5


Disassembly of the medTester 5000C’s Feet and Tilt Bail
Disassemble the medTester 5000C feet and tilt bail. Prior to disassembly,
clear the work area of any objects that could scratch or puncture the top of
the medTester. Cover the work area with a cloth.

1.

Turn the power off to the medTester.

2.

Unplug the medTester from the power outlet.

3.

Lay the medTester face down on the work area.

4.

Using a #2-type Phillips screwdriver, unscrew and remove the
screws holding the medTester feet and tilt bail. Save the screws,
feet, and tilt bail for possible future use.

Warning!

The tilt bail is under tension.

Attaching the Wedge to the medTester 5000C
Attach the Wedge to the bottom of the medTester 5000C and connect the
required cables:

1.

Disassemble the two halves of the Wedge case.

2.

Save the six screws, three from the front half, three from the rear
half.

3.

Lay the top cover of the Wedge on the bottom of the medTester,
with the thick end of the Wedge toward the rear, and line up the
screw holes on the Wedge to the screw holes on the medTester.

4.

Using the four supplied screws, screw the Wedge top cover to the
bottom of the medTester.

5.

Place the bottom half of the Wedge, which contains the circuit
board, into the Wedge cover.

6.

Reassemble the Wedge by screwing together the two halves of the
case using the original screws.

7.

Connect the two Wedge cables from the Wedge to the
medTester 5000C as follows: