Biomedical Test System

General Information

1

1-15

Ground Fault Circuit Interrupter

The test receptacle is protected by a ground fault circuit interrupter. In the
event of a fault in the EUT (when the test receptacle is powered on) which
causes a current imbalance in the hot and neutral lines of greater than 5 mA,
the interrupter triggers. Under this condition, the interrupter turns off the test
receptacle and all test connections, and the medTester prompts you to correct
the fault and continue.
Performance Waveforms
The medTester 5000C generates ECG waves and arrhythmias to test the
performance of patient monitors in those medTesters that have the
Waveform/Extended Testing Module installed. You can find information
about conducting waveform tests in Chapter 7, Performance Waves.
Specifications for waveforms and arrhythmias are:

ECG Performance Test Waves (Lead I, Vp-p)

Square Wave—2 Hz, 1 mV

DC Pulse—4 Seconds, 1 mV

Sine Wave—0.5, 10, 40, 60, and 100 Hz, 1 mV

Square Wave—1 kHz, 1 mV

Triangle—2 Hz, 1 mV

CMRR—60-Hz sine wave with 1–k

Ω imbalance in LA

Normal Sinus—30, 60, 120, 240 BPM.

Arrhythmias

Atrial fibrillation

Second degree A-V Block, Type 1

Premature atrial contractions

Missed beat at 80 and 120 BPM

PVC 1 left

PVC 2 right

Multifocal PVCs

PVC 1, R on T

A pair of PVCs

Run of 5 PVCs

Run of 11 PVCs, MF

Right bundle branch block

Ventricular tachycardia

Ventricular fibrillation

Asystole