Biomedical Test System

Defibrillator Module

9

9-7

Note

See Chapter 5, Autosequences, for a complete explanation of
working with prompts. You can turn individual prompts on or
off and rename prompts six through eight (6-8) from the list
above. Chapter 6, Customize Your medTester 5000C, explains how
to customize prompts.

Test Sequence
After you input pretest prompt information, defibrillator tests execute in the
order described below.

Plug the defibrillator (the EUT, equipment under test) into an appropriate
power source.

Note

Since defibrillators may produce electrical noise that could
interfere with test measurements on the medTester 5000C, it is
advised to plug the defibrillator into another power receptacle.
However, during any paddle tests, the defibrillator must be
plugged into the medTester 5000C’s test receptacle.

APEX

Paddle.

The APEX paddle test performs two (2) leakage measurements.

One with closed ground; the other with open ground. Both measurements are
taken under normal polarity and closed neutral conditions. Before these tests
execute, this message displays:

Press

READY

to begin the tests. Measurements are in µA RMS.

Connect EXT red probe to APEX paddle

READY

F5