Fluke Biomedical medTester 5000C User Manual

Page 192

Advertising
background image

medTester 5000C
Operators Manual

12-14

MEASUREMENT COMMANDS

Note: These commands set up a measurement, according to global variables as needed, and
after a 0.5 second settling time, return the measurement reading. The measurement remains set
up, so that additional readings can be taken with the REPEAT command. Global variables can
be changed while the measurement remains set up, then more readings can again be taken
with the REPEAT command. A measurement can be turned off by any non-measurement/non-
setup command or by the EXIT command. While a measurement is set up, a new measurement
command is allowed without previously exiting. The units of measurement are listed in
parentheses. For ECG lead isolation tests, the isolation voltage stays on only during each
measurement, then turns off. A REPEAT command would turn it back on as long as needed for
the next measurement.

Command

Explanation

The response is indicated with the unit of measurement listed in

parentheses.

SV12

System voltage, hot to neutral.
(V RMS)

SV1G

System voltage, neutral to ground.
(V RMS)

SV2G

System voltage, hot to ground.
(V RMS)

CLKE

Case leakage current, external leads.
(µA RMS or DC)

CLKI

Case leakage current, internal leads.
(µA RMS or DC)

XTMI

External meter leakage current.
(µA)

ALGD

ECG leakage current, All leads to ground.
(µA RMS or DC)

RLGD

ECG leakage current, Right Leg to ground.
(µA RMS or DC)

RAGD

ECG leakage current, Right Arm to ground.
(µA RMS or DC)

LAGD

ECG leakage current, Left Arm to ground.
(µA RMS or DC)

LLGD

ECG leakage current, Left Leg to ground.
(µA RMS or DC)

VGND

ECG leakage current, V leads to ground.
(µA RMS or DC)

RLLD

ECG leakage current, Right Leg to leads.
(µA RMS or DC)

RALD

ECG leakage current, Right Arm to leads.