Temperature, Humidity – Fluke Biomedical medTester 5000C User Manual

Page 54

Advertising
background image

medTester 5000C
Operators Manual

4-8

4.

Select the keyboard function key that corresponds to the lead you
wish to test.

5.

To begin the measurement, press

ISOV ON

F5

. The test begins,

and the

ISO VOLTS

annunciator lights. The reading is displayed for

up to 15 seconds. You can escape to terminate the test. The test
self-terminates after 15 seconds.

Note

If you receive abnormally high leakage current measurements
with V1-V6 leads tied together, you can inspect individual V
leads by removing the other leads. This procedure allows you to
isolate the lead with a high leakage current reading.

Equipment Current
See Figure 4-8, Equipment Load Current.

This test measures current used by the equipment under test. The EUT is
plugged into the

TEST RECEPTACLE

. The receptacle is fixed at normal polarity,

closed ground, and closed neutral. The range of this test is 20 amps RMS with
an accuracy of

±5% of the range.

1.

Plug the EUT into the

TEST RECEPTACLE

.

2.

Enter menu commands:

MANUAL

F2

EQU CUR

F3

Resistance
These two resistance tests use the four-terminal test method. Resistance is
measured in a range of 2.000

Ω with an accuracy of ±1% of range with a test

current of 100 mA.