Operating the wedge, Wedge port names, Enabling the wedge – Fluke Biomedical medTester 5000C User Manual

Page 199: Configuring wedge ports

Advertising
background image

Biomedical Test System

The Wedge Adapter

13

13-3

When you execute a checklist, the medTester 5000C directs any M-type
commands through the device connected to the checklist port. If the
command requires a measurement response, for example, that input returns
to the medTester through the checklist port as part of the test record. In
addition, one or more ports can be used for autosequences. Defibrillator
autosequences can execute through one port you define, IV pump
autosequences through another port, and so on.

The table below describes the types of port configurations, their default port
settings, and descriptions.

Wedge Port Configurations

Wedge Port

Name

(Default Wedge

Port No.)

Description

Normal

(1)

Input in Local Mode—laser bar code scanning gun or a personal computer or terminal.

Output in Local Mode—for printing test records to a personal computer or compatible terminal device,
or test tags to a serial printer.

Two-way Communications—medTester control from a remote personal computer or a compatible
terminal device.

See Chapter 12, Remote Operation for more information.

Defibrillator

Autosequences

(2)

Execute medTester 5000C autosequences A30-A49.

Specific to Impulse Models 4000 and 3000 Defibrillator Analyzers.

Both preprogrammed and user-configured autosequence types.

See Chapter 9, Defibrillator Module, for more information.

External Pacemaker

Tester (PACER)

Autosequences

(2)

Execute medTester 5000C PACER autosequences.

Compatabile with the Impulse 4000 Analyzer Transcutaneous pacemaker testing features.

Both preprogrammed and user-configured autosequence types.

See Chapter 16, Transcutaneous Pacemaker (Pacer) Module, for more information.

IV Pump

Autosequences

(3)

Execute medTester 5000C autosequences A50-A59.

Compatable with the IDA 4, Infutest 2000 series D, IPT-MC, and the IPT-I IV Pump Analyzers.

Both preprogrammed and user-configured autosequence types.

See Chapter 10, IV Pump Module, for more information.

Electrosurgical Unit

(ESU)

Autosequences

(4)

Execute medTester 5000C ESU autosequences.

Compatabile with the Models 454A, RF303, and 402A ESU Analyzers.

Both preprogrammed and user-configured autosequence types.

See Chapter 14, ESU Module, for more information.

Pulse Oximetry

(SpO2)

Autosequences

(5)

Execute medTester 5000C SpO2 autosequences.

Compatible with the models Index 2XL, Oxitest PLUS, Oxitest PLUS7, and the CardioSat100 series
SPO2 simulator.

Both preprogrammed and user-configured autosequence types.

See Chapter 15, SPO2 Module, for more information.