Esu autosequence descriptions – Fluke Biomedical medTester 5000C User Manual

Page 216

Advertising
background image

medTester 5000C
Operators Manual

14-6

To accept previously input prompt information, press

YES

, and go to the Test

Sequence section. If you choose

NO

, go to the Pretest Device Prompts section that

follows.

Pretest Device Prompts

There are eight prompts for information that can attach to test records. After
selecting the desired autosequence, the medTester 5000C prompts you to
enter device prompts. If you have some or all prompts turned off, the
autosequence still gives you the brief opportunity to enter device information
at the tests initiation with the message:

1.

OP CODE—This is the operator code for the person doing the
testing. This code can consist of a maximum of three characters
which can be entered from the medTester keyboard. An example
of an operator code could be a person’s initials or employee
number. The previous operator’s code, if any, will appear. If the
current operator has not changed from previous use, press the

ENT

key.

2.

TYPE—Enter the device name, code, or other descriptor for the
equipment under test. This entry can be a maximum of 16
characters.

3.

MANF—Enter the name of the device manufacturer, 16
characters maximum.

4.

LOC—Type in the location of the device, 16 characters
maximum. The previous location, if any, will appear. If this occurs
and is the desired location, press the

ENT

key.

5.

MODEL—Enter the device model number, 16 characters
maximum.

6.

SERIAL #—Type the device serial number, again 16 characters
maximum.

7.

CONTROL #—Enter a control number with a maximum of 16
characters.

Press F5 for all device info prompts