Biomedical Test System

ESU Module

14

14-9

CQM (Contact Quality Monitor) Test

The CQM test measures the resistance needed by the ESU device to activate
the CQM alarm. For this test the 454A requires an auxiliary module, either
the REM/ARM module or the combined REM/ARM and Return Fault
module (RARF).

Press

F5 READY

to start the test. Connect the ESU analyzer and the ESU

device for a CQM (REM/ARM) test. On the 454A and 402, adjust the
resistance on the module until the ESU alarm is activated. On the RF303

RS

, a

CQM test uses the analyzer’s main test load and defaults to 100 ohms. You
can adjust this level when it is displayed on the RF303

RS

during a CQM test

by pressing the INCREASE/DECREASE TEST LOAD keys on the
RF303

RS

front panel until the desired resistance value is displayed. Press

F5

READ

to read the resistance on the medTester. You can accept this reading

and store PASS or FAIL by pressing

F4 PASS

or

F5 FAIL

respectively. FAIL

stores a "#" with the test record. You can press

F3 REPEAT

to go back to the

measurement screen.

Return Fault Test (for Model 454A only)

This test checks the return fault monitor on an ESU. For this test the 454A
requires the EM/ARM and Return Fault module (RARF). The setup screen
prompts the user to set up the ESU for power setting and to select either the
cut or coag side of the ESU device.

Press

F5 READY

to start. Connect the ESU to the ESU analyzer for a return

fault monitor test. First, the ESU analyzer sets up the No Fault condition.
The ESU should not alarm. Press

F4 PASS

or

F5 FAIL

to store the test results.

As before, FAIL stores a "#" along with it. Second, the ESU analyzer sets up
the Fault condition. The ESU should alarm. Press

F4 PASS

or

F5 FAIL

to store

the test results.

The table beginning on the next page supplies specific information about the
tests for each autosequence.