medTester 5000C
Operators Manual

15-26

MORE

F4

to select more steps, or

END

F5

to end your selections.

6.

STARTING ALARM VALUES: Enter starting values for the alarm
tests for:

SPO2

HEART RATE

AMPLITUDE (For the Index 2XL, enter amplitude from 0% to
20% with the thirty-three selections offered)

Note

Even if you do not intend to perform any alarm tests, you still
have to go through the starting alarm values. You can just press

ENT

to get through them.

7.

Answer F4 YES or F5 NO whether to perform the following special tests:

SPO2 ALARM TEST: If YES, then enter the Hi and Lo alarm
values.

RATE ALARM TEST: If YES, then enter the Hi and Lo alarm
values.

ASYSTOLE TEST

AMPLITUDE TEST

MOTION TEST (Index 2XL only) – Select the motion test desired.

PROBE TEST

After customizing your autosequence, the medTester prompts you to store
the autosequence:

STORE

F5