Using the ESU Module with the Model 402A and RF303

RS

ESU Analyzer

The ESU module was originally designed to work with the Fluke
Biomedical Model 454A ESU Analyzer. Fluke Biomedical’s
Model 402A and RF303

RS

ESU Analyzers are also compatible

with the medTester, but operate differently sometimes from the
Model 454A. The medTester can run ESU autosequences with
these analyzers but must be customized for the capabilities of
the ESU Analyzer connected. The Model 402A and RF303

RS

both have similar functionality. They have fewer test loads
available, no plug-in modules, and connects for its test
differently. They can not perform a return fault test. If an
autosequence attempts to run a non-compatible test, it will not
execute and the entire autosequence will be aborted.
For further information on using the Model 402A with the
medTester, please refer to Chapter 6, Remote Operation: medTester
5000C Autosequence Operation
, in the Model 402A Operating
Manual. Remember that you must use autosequences that are
within the capability of the Model 402A.

When you connect a Fluke Biomedical ESU analyzer to your medTester and
run ESU autosequences, the medTester displays prompts for your ESU setup
and testing and test execution. The ESU analyzer returns measurement
readings to the medTester, which displays them for you. If you have the
medTester 5000C RS-232/Printer module and the Memory module, you can
store and print test records.

Having the ESU module also offers you the ability to incorporate these
autosequences into checklists if your medTester 5000C has the medCheck
module installed.

For related information, see these chapters in this manual:

Chapter 5, Autosequences.

Chapter 6, Customize Your medTester 5000C—for customizing
autosequences.

Chapter 8, Memory—for test record storage and printing.