Setting output ports – Fluke Biomedical medTester 5000C User Manual

Page 205

Advertising
background image

Biomedical Test System

The Wedge Adapter

13

13-9

Configuring an Autosequence Port

You can execute defibrillator, IV pump, ESU, and SPO2 autosequences
through a Wedge port in the same manner that you can execute them through
the medTester 5000C’s COM2 port. This allows you, for example, to connect
both a defibrillator analyzer such as the Impulse 3000 or the Impulse 4000
and an IV pump tester such as the IDA 4 to defined ports on the Wedge.
Then whenever you execute one of these autosequences, the
medTester 5000C knows to which port it must send the autosequence.

The medTester 5000C also uses the configured autosequence port for
executing any autosequence that is embedded as a checkitem in a checklist.
For example, if the medTester executes a checklist that contains a defibrillator
autosequence, it directs that autosequence to the defibrillator autosequence
port.

Each autosequence type has a default port setting. Defibrillator autosequences
are assigned to the Wedge’s COM2-2 port. Follow these instructions to
configure a different port for defibrillator autosequences:

1.

From the medTester 5000C main menu,

MENU 1

, press:

UTIL

F5

2.

From the utility menu, press the medTester right arrow and enter
the command:

WPORTS

F5

3.

Press:

DEFIB

F2

to see this display:

4.

Press the medTester up arrow key to increment the number to the
port you want as the defibrillator autosequence port, or press the
down arrow key to decrement the number.

5.

After selecting the normal port, press

F5

to store your selection.

SET WEDGE PORT FOR DEFIB: 2

STORE

F5