Print your autosequence, User output names, Reinitialize factory default settings – Fluke Biomedical medTester 5000C User Manual

Page 224: Comments, Next test due date, User time, User time is required, Print test tag, Store or print test record

Advertising
background image

medTester 5000C
Operators Manual

14-14

Autosequence Number: A64

Autosequence Name: VLSSE4

Step and Test

Type

Load

ESU Mode

Output)

Power

Setting

Expected Power

or

Electrode

Limits

1

Generator Output

300 Ohms

Monocoag

Spray

25

25 W

± 20%

2

Generator Output

300 Ohms

Monocoag

Spray

120

120 W

± 15%

3

Generator Output

300 Ohms

Monocoag

Soft

60

60 W

± 15%

4

Generator Output

300 Ohms

Monopolar

Pure Cut

75

75 W

± 15%

5

Generator Output

300 Ohms

Monopolar

Pure Cut

300

300 W

± 15%

6

Generator Output

300 Ohms

Monopolar

Blend 1

200

200 W

± 15%

7

Generator Output

300 Ohms

Monopolar

Blend 2

200

200 W

± 15%

8

Generator Output

300 Ohms

Monopolar

Blend 3

200

200 W

± 15%

9

Generator Output

300 Ohms

Monopolar

Blend 1

75

75 W

± 15%

10

Generator Output

300 Ohms

Monopolar

Blend 2

75

75 W

± 15%

11

Generator Output

300 Ohms

Monopolar

Blend 3

75

75 W

± 15%

12

Generator Output

100 Ohms

Bicoag

Standard

70

70 W

± 15%

13

Generator Output

100 Ohms

Bicoag

Precise

70

70 W

± 15%

Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:

1.

COMMENTS

2.

NEXT TEST DUE DATE

3.

USER TIME

EMS/CMMS system users:

User time is required

4.

PRINT TEST TAG

5.

STORE or PRINT TEST RECORD