ESU Autosequence Descriptions

Autosequence Number: A60

Autosequence Name: ASPENX

Step and Test

Type

Load

ESU Mode

Output

Power

Setting

Expected Power

or

Electrode

Limits

1

Generator Output

300 Ohms

Monopolar

Pure Cut

120

120 W

± 15%

2

Generator Output

300 Ohms

Monopolar

Pure Cut

300

300 W

± 15%

3

Generator Output

300 Ohms

Monopolar

Blend 1

69

69 W

± 15%

4

Generator Output

300 Ohms

Monopolar

Blend 1

180

180 W

± 15%

5

Generator Output

300 Ohms

Monocoag

Standard

48

48 W

± 15%

6

Generator Output

300 Ohms

Monocoag

Standard

120

120 W

± 15%

7

Generator Output

300 Ohms

Monocoag

Spray

40

40 W

± 15%

8

Generator Output

300 Ohms

Monocoag

Spray

80

80 W

± 15%

9

RF Leakage

Open Monocoag Standard 120 Active

Electrode N/A

10

RF Leakage

Open Monocoag Standard 120

Dispersive

Electrode

N/A

11

RF Leakage

Open

Monopolar

Pure Cut

300

Active Electrode

N/A

12

RF Leakage

Open Monopolar Pure

Cut 300

Dispersive

Electrode

N/A

13

Generator Output

50 Ohms

Bipolar

Coag

20

20 W

± 20%

14

Generator Output

50 Ohms

Bipolar

Coag

50

50 W

± 15%

15

Generator Output

50 Ohms

Bipolar

Cut

20

20 W

± 20%

16

Generator Output

50 Ohms

Bipolar

Cut

50

50 W

± 15%