medTester 5000C
Operators Manual

9-14

Reinitialize Factory Default Settings

Any customized autosequences can be returned to the factory default settings.
To do this, enter these medTester commands:

1.

From the medTester 5000C main menu,

MENU

1, press the

right arrow to access

MENU2

.

2.

Enter the command:

CUSTOM

F3

3.

From the custom menu, press the right arrow and enter the

command:

INIT

F3

4.

Choose:

DEFIB

F2

5.

To initialize defibrillator autosequences to defaults, press:

YES

F5

Note

Reinitializing will return all customized or created autosequences
to factory default settings, including three blanks.