Fluke Biomedical medTester 5000C User Manual
Page 138
Advertising
medTester 5000C
Operators Manual
9-14
Reinitialize Factory Default Settings
Any customized autosequences can be returned to the factory default settings.
To do this, enter these medTester commands:
1.
From the medTester 5000C main menu,
MENU
1, press the
right arrow to access
MENU2
.
2.
Enter the command:
CUSTOM
F3
3.
From the custom menu, press the right arrow and enter the
command:
INIT
F3
4.
Choose:
DEFIB
F2
5.
To initialize defibrillator autosequences to defaults, press:
YES
F5
Note
Reinitializing will return all customized or created autosequences
to factory default settings, including three blanks.