Running defibrillator autosequence tests – Fluke Biomedical medTester 5000C User Manual

Page 129

Advertising
background image

Biomedical Test System

Defibrillator Module

9

9-5

Running Defibrillator Autosequence Tests

To begin running defibrillator autosequence tests, connect your
medTester 5000C to your Impulse 4000 or Impulse 3000. These two
instruments are connected through the COM2 port on the medTester and the
RS-232 serial port on the Impulse model you have.

Note

Refer to Chapter 1, General Information- Accessories , or the current
Fluke Biomedical Price List for the correct serial cable part
number needed.

From the medTester 5000C defibrillator menu, select the autosequence name
for the autosequence you wish to run. The name and number of the selected
autosequence appears on the display followed by the prompt:

If a safety autosequence has executed since power up of the medTester, you
can have the medTester use the previously input prompt information for the
current defibrillator autosequence, from this prompt:

To accept previously input prompt information, press

YES

, and go to Test

Sequence, which follows this section. If you choose

NO

, go to the prompt

instructions in Pretest Device Prompts.

Press F5 for all device info prompts

USE PREVIOUS STARTING INFORMATION?