For the oxitest plus – Fluke Biomedical medTester 5000C User Manual

Page 242

Advertising
background image

medTester 5000C
Operators Manual

15-14

For the Oxitest PLUS:

SPO2 Autosequence Number: A72

Autosequence Name: ADULT3

Settling Time: 25 Beats

Heart Rate Limit:

± 4 BPM (beats per minute)

Step

Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Artifact

1 97 ± 4

75 100

Off

2 97 ± 4

45 100

Off

3 97 ± 4

180 100

Off

4 80 ± 4

45 100

Off

5 90 ± 4

120 100

Off

6 80 ± 4

180 100

Off

7 97 ± 4

66

4

Off

8 97 ± 4

70 33

Movement

9 97 ± 4

80 52

Tremor

Starting Alarm Value for SPO2: 93%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 100%
SPO2 Alarm Test: Hi–97% Lo–80%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test

SPO2 Autosequence Number: A73

Autosequence Name: NEO1

Settling Time: 25 Beats

Heart Rate Limit:

± 6 BPM (beats per minute)

Step

Number

SPO2

(in percent)

SPO2 Limit

(in percent)

Heart Rate

(in BPM)

Amplitude

(in percent)

Artifact

1 93 ± 4

145

40

Off

2 93 ± 4

75

40

Off

3 93 ± 4

225

40

Off

4 97 ± 4

225

40

Off

5 80 ± 4

60

40

Off

6 93 ± 4

145

40

Movement

7 93 ± 4

145

40

Tremor