medTester 5000C
Operators Manual

17-10

NIBP Autosequence Number: A80

Autosequence Name: ADULT1

Step Number

Blood Pressure

Heart Rate in BPM
(beats per minute)

Envelope Gain

Envelope Shift

1

120/80 (90) ADULT

80

100%

+0

2

80/50 (62) ADULT

45

100%

+0

3

150/100 (115) ADULT

125

100%

+0

NIBP Test Limits:

NIBP: ± 20%

Heart Rate: ± 4 BPM

Pressure Test: No
Leak Test: No
Pop Off Test: No

NIBP Autosequence Number: A81

Autosequence Name: ADULT2

Step Number

Blood Pressure

Heart Rate in BPM
(beats per minute)

Envelope Gain

Envelope Shift

1

120/80 (90) ADULT

80

100%

+0

2

80/50 (62) ADULT

45

100%

+0

3

150/100 (115) ADULT

125

100%

+0

NIBP Test Limits:

NIBP: ± 20%

Heart Rate: ± 4 BPM

Pressure Test: Yes
Leak Test: Yes
Pop Off Test: Yes

NIBP Autosequence Number: A82

Autosequence Name: ADULT3

Step Number

Blood Pressure

Heart Rate in BPM
(beats per minute)

Envelope Gain

Envelope Shift

1

120/80 (90) ADULT

80

100%

+0

2

80/50 (62) ADULT

80

100%

+0

3

150/100 (115) ADULT

80

100%

+0

4

120/80 (90) ADULT

45

100%

+0

5

120/80 (90) ADULT

125

100%

+0

6

200/150 (165) ADULT

200

100%

+0

7

120/80 (90) AFIB

N/A

N/A

N/A

8

120/80 (90) PAC

N/A

N/A

N/A

NIBP Test Limits:

NIBP: ± 20%

Heart Rate: ± 4 BPM

Pressure Test: Yes
Leak Test: Yes
Pop Off Test: Yes